Recall Z-0652-2009 ยท HRX Arthroscope

Stryker Endoscopy

Stryker Endoscopy FloControl Arthroscopy Pump Integrated Tubing, Model number 0350600006, for use with the FloControl Arthroscopy Pump.

DateJan 21, 2009 (initiated Oct 1, 2008)
StatusTerminated
ReasonUnable to pump fluids. Silicone tubing is stiffer than usual, resulting in motor defect message, leading to inability to pump saline through tubes for surgery.
Root causeProcess control
ActionStryker Endoscopy initiated the recall 10/01/2008. Information gathered and recall packets sent via FedEx to all affected accounts. Packets included advisory notices and acknowledgement card, as well as shipping labels to return product. Each account was instructed to locate and quarantine product for return to Stryker endoscopy. Sales representatives received a voicemail/email notification that included a list of all affected accounts.
Quantity1440 devices
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K944210ARTHROSCOPE ACCESSORIES
Stryker Endoscopy, Feb 14, 1995