Recall Z-0001-2009 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Philips Medical Systems North America Co. Phillips

NeuViz Dual Multi-Slice Computed Tomography (CT) Scanner System (software). Whole body x-ray computed tomography scanner which features a continuously rotating tube-detector system and functions according to the fan beam principle. The system software is an interactive program used for x-ray scan control, image reconstruction, and image archive/evaluation. It is intended to produce cross-section

DateJan 26, 2009 (initiated Apr 10, 2008)
StatusTerminated
ReasonAn issue associated with the "Timed Scan" application was found. The system does not wait for the programmed time interval and starts the x-ray after selection of the scan button when certain criteria is met.
Root causeSoftware design
ActionA letter was issued to customers prior to the release of the field correction during June 2008. The system software will be updated to correct the non-compliances. For information or support concerning this issue, please contact the Neusoft Medical Systems Service Support Department (helpdesk@pnms.neusoft.com) or contact Philips Medical Systems North America Co. Phillips at 425-487-7602.
Quantity19 units in the US.
DistributionNationwide Distribution --- USA including states of CO, FL, GA, IL, KY, NC, TN and TX.
FDA recordOpen on fda.gov