ClearanceBrief
Recall Z-0287-2010 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems (Cleveland) Inc

Respiratory Gating System, Catalog Number: NPTA540, Model #4535 674 26701, for the following affected CT/PET scanning device systems : GEMINI TF CT/PET Systems 16 Slice, GEMINI TF CT/PET System 16 Slice Mobile; GEMINI TF CT/PET System 64 Slice Mobile; and GEMINI TF CT/PET System 64 Slice. (Model Numbers: 16 slice: 4535 679 83931; 64 slice Mobile: 4535 674 47551; and 64 slice: 4535 679 94741.

DateNov 17, 2009 (initiated Nov 24, 2008)
StatusTerminated
ReasonA leak between the Tube Interface and the Outlet Tube of the transducer for the Respiratory Gating System component may result in a failure of the device to produce respiratory correlated images.
Root causeProcess design
ActionPhilips Medical Systems Nuclear Medicine division initiated this recall by sending a 'Customer Advisory Notification to their customer accounts on 11/24/2008. This notification informs the customer of the potential for the device equipped with the Pulmonary Gating Option to develop an air leak which could result in the device being unable to produce the desired respiratory correlated images. The notification provides field test instructions and recommends that the customer test the bellows on each day of use in order to assess that the unit is functioning properly. If the bellows device fail
Quantity51 units
DistributionWorldwide distribution: USA, Puerto Rico, Canada, Belgium, China, Denmark, Egypt, France, Germany, Hong Kong, India, Japan, Netherlands, Pakistan, Philippines, Poland, Saudi Arabia, Singapore, South Korea, Spain, and Sweden.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033357ACQSIM-MULTISLICE-CT, MODEL 453567042271
Philips Medical Systems (Cleveland), Inc., Jan 9, 2004
K052640GEMINI RAPTOR
Philips Medical Systems (Cleveland), Inc., Oct 7, 2005