Recall Z-1091-2009 ยท CGA Glucose Oxidase, Glucose

Pointe Scientific, Inc.

Glucose Oxidase Reagent as follows: Pointe Scientific Glucose Oxidase Reagent Set, Pointe Scientific, Inc, Canton, MI; Catalog # G7519-1L, G7519-500 and G7519-SAM. 2MM Diagnostics Glucose Oxidase Reagent Set, 2MM Corporation, Miami, FL; Reorder # RT-160-500. Pointe Scientific Glucose Oxidase Reagent, Pointe Scientific, Inc., Canton, MI; Catalog, MI; Catalog # 2-G7519-L, 7-G7519-100, 7-G7519-20

DateMar 12, 2009 (initiated Nov 3, 2008)
StatusTerminated
ReasonThe product is unable to maintain the specification for linearity through the shelf life of the product.
Root causeProcess control
ActionConsignees were notified via an Urgent: Medical Device Recall letter dated 11/3/08 that all test results obtained with these lots are questionable and that attending physicians should be informed to determine whether additional testing is required. Also, consignees should examine their inventories for any of the lot numbers listed and dispose of them. If the product has been further distributed, identify customers and notify them of the product recall by including a copy of the recall letter. The Recall Return Response Form should be completed and returned and upon receipt Pointe Scientific
Quantity1500 liters
DistributionWorldwide Distribution -- USA, Hong Kong, Indonesia, Israel, Jordan, Mexico, Nigeria, Poland, Russia, Serbia, Sri Lanka, Trinidad, and Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K810674GLUCOSE OXIDASE
Pointe Scientific, Inc.,, Mar 31, 1981