Recall Z-0327-2009 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Siemens Medical Solutions USA, Inc

Axiom Luminos dRF without ST filter. . Solid state x-ray imager (flat panel/digital imager). Model number 10252048. Universal fluoroscopic x-ray diagnostic system.

DateNov 28, 2008 (initiated Sep 19, 2008)
StatusTerminated
ReasonIssue may occur where an intended movement of the joystick may unintentionally initiate an x-ray exposure.
Root causeNonconforming Material/Component
ActionThe recalling firm issued Update Instruction AX042/08/S dated August 2008, to affected customers. This Update instruction provided Siemens service engineers with instructions for the replacement of the joystick module at each facility.
Quantity1 unit
DistributionNationwide Distribution --- including states of KY, NJ and PA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062623AXIOM LUMINOS DRF
Siemens Medical Solutions USA, Inc., Aug 22, 2007