Recall Z-0597-2009 ยท KPS System, Tomography, Computed, Emission

Philips Medical Systems (Cleveland) Inc

Vertical drive brake (on patient supports). Used on certain Philips Medical Systems MX series and Brilliance series CT scanners. The device is a vertical drive brake that controls the vertical position of a patient support system or couch. The patient support system is intended to be used to place the patient in the correct horizontal and vertical position in order to be scanned. The couch is a

DateJan 16, 2009 (initiated Sep 23, 2008)
StatusTerminated
ReasonThe product is being recalled because the patient support may travel downwards without being commanded to move.
Root causeProcess control
ActionA On 11/15/2007 & 6/6/2008 the firm sent Urgent Device Notification letters to their customers. They notified their customers that a Field Service Engineer will inspect and repair the brake system.
Quantity559
DistributionWorldwide Distribution: USA including states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and WV and countries of Argentina, Australia, Austria, Bahamas, Belgium, Brazil, Can
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010817MX8000 V5.0 CT SYSTEM
Philips Medical Systems (Cleveland), Inc., Jun 15, 2001
K012009MX8000 IDT CT SYSTEM
Marconi Medical Systems, Inc., Sep 13, 2001
K033326PHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM
Philips Medical Systems, Inc., Oct 29, 2003
K033357ACQSIM-MULTISLICE-CT, MODEL 453567042271
Philips Medical Systems (Cleveland), Inc., Jan 9, 2004