ClearanceBrief
Recall Z-2477-2010 ยท IZI System, X-Ray, Angiographic

Philips Medical Systems North America Co. Phillips

Philips Integris Allura Flat Detector 9C Angiographic X-ray system. This system with Velara Generator can be identified as type number 9890 000 PRODUCTS 6501 * (number located on the generator cabinet). Intended use for Angiographic X-ray system with and without a solid state X-ray imaging device for cardiovascular and vascular diagnostic and interventional procedures.

DateSep 22, 2010 (initiated Oct 23, 2008)
StatusTerminated
ReasonA high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the display.
Root causeOther
ActionThe firm, Philips, sent "URGENT-Device Correction" letters dated November 10, 2008, to their customers and informed them about the reliability issue in the Allura XPER and Integris X-ray systems with the High Voltage Velera (HV) Generator. This reliability issue can lead to a system error causing intermittent loss of fluro and/or exposure acquisition, potentially resulting in the loss of the X-ray image on the display. The customers are advised if they experience High Voltage Generator reliability issue and the HV Generator stops working generating an O3EA error, a cold restart is required a
Quantity269 units in US
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022899PHILIPS INTEGRIS ALLURA FLAT DETECTOR RELEASE 1.0
Philips Medical Systems North America Co., Nov 22, 2002