Recall Z-2474-2010 ยท IZI System, X-Ray, Angiographic
Philips Medical Systems North America Co. Phillips
Philips Integris XPER FD10 F/C Angiographic X-ray system. This system with Velara Generator can be identified as type number 9890 000 6501 * (number located on the generator cabinet). Intended use for Angiographic X-ray system with and without a solid state X-ray imaging device for cardiovascular and vascular diagnostic and interventional procedures.
| Date | Sep 22, 2010 (initiated Oct 23, 2008) |
| Status | Terminated |
| Reason | A high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the display. |
| Root cause | Other |
| Action | The firm, Philips, sent "URGENT-Device Correction" letters dated November 10, 2008, to their customers and informed them about the reliability issue in the Allura XPER and Integris X-ray systems with the High Voltage Velera (HV) Generator. This reliability issue can lead to a system error causing intermittent loss of fluro and/or exposure acquisition, potentially resulting in the loss of the X-ray image on the display. The customers are advised if they experience High Voltage Generator reliability issue and the HV Generator stops working generating an O3EA error, a cold restart is required a |
| Quantity | 206 units in US |
| Distribution | Worldwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K022899 | PHILIPS INTEGRIS ALLURA FLAT DETECTOR RELEASE 1.0 Philips Medical Systems North America Co., Nov 22, 2002 |