ClearanceBrief
Recall Z-0330-2009 ยท FCG Biopsy Needle

Boston Scientific Corporation

Boston Scientific Easy Core" Biopsy System, 15 ga. x 15 cm, REF 43-451, UPN M001434511, Sterilized with ethylene oxide gas. UPN for inner pouch M001434510. Indicated for use endoscopically or percutaneously to retrieve tissue sampling of soft organs, tumors or masses for histological analysis. Soft tissue sampling includes, but is not limited to, organs such as breast, liver, kidney or prostat

DateNov 28, 2008 (initiated Oct 7, 2008)
StatusTerminated
ReasonBoston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking or arming the cannula latch on the device. This difficulty may result in an inability to use the device.
Root causeDevice Design
ActionConsignees were sent a "Boston Scientific Urgent Medical Device Recall" letter dated October 21, 2008. The letter was addressed to The Hospital Field Action Contact. The letter described the problem and products being recalled. The letter stated that it was very important that the attached product table be used carefully and consider both inner and outer packaging product codes when searching for affected/recalled product as the UPN numbers on the inner and outer labeling are different. Requested to discontinue use of product, identify, and segregate and return affected products to BSC.
Quantity3808 total for all products
DistributionNationwide Distribution --- including states of AL, AZ, AR, CA, CT, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO,MT, NE , NV, NH, NH, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, and PR.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040893EASY CORE BIOPSY SYSTEM
Boston Scientific Corp, May 6, 2004