Recall Z-0414-2009 ยท BTT Humidifier, Respiratory Gas, (Direct Patient Interface)

Fisher & Paykel Healthcare Inc

Fisher & Paykel Healthcare RT240 Adult Breathing Circuit Kit. lntended Use: The dual heated adult respiratory ventilator circuits are intended as conduits of breathing gas for ventilation of patients, and to maintain the temperature of the humidified gas, to reduce condensation.

DateDec 10, 2008 (initiated Oct 22, 2008)
StatusTerminated
ReasonThe recall was initiated because certain lot dates of the RT240 Adult Breathing Circuit Kits manufactured on or before June 27, 2008 include a heated breathing circuit that may be more susceptible to damage when used in excess of the specified seven-day maximum duration of use, which may increase the risk of malfunction or fire.
Root causeProcess change control
ActionFisher & Paykel Healthcare, Inc. (FPH) issued an URGENT-PRODUCT RECALL letter (dated October 8, 2008) including a Customer recall letter for consignees/customers to Cardinal Health. Cardinal Health will, in turn, send the Customer recall letter dated October 10, 2008 to its consignees/customers. Fisher & Paykel Healthcare (FPH) also issued a similar set of recall notification letters to Norco on October 15, 2008. Recipients of the recall letter must perform the following: 1) Check all Cardinal Health stock of the product for the affected lots, 2) Immediately quarantine the affected product
Quantity229,320 units
DistributionWorldwide Distribution including US and country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K983112RESPIRATORY HUMIDIFIER, MODEL # MR850
Fisher & Paykel Healthcare, Ltd., Nov 10, 1998