Recall Z-0412-2009 ยท FSA Lift, Patient, Non-Ac-Powered

Arjo, Inc.

Arjo Maxi 500 Patient Lift with 2-Point Spreader Bar and Scale. A Non-AC-Powered Patient Lift. BHM Medical Inc., Magog, QC, Canada. Model KM560101, and Model KM560001 - Maxi 500 with 2-Point Spreader Bar, if it was equipped with the optional scale kit 700.05505. The Maxi 500 is a mobile passive lifter, intended to be used for lifting and transferring of patients in hospitals, nursing homes of

DateDec 18, 2008 (initiated Oct 30, 2008)
StatusTerminated
ReasonIf the spring pin is not properly reinstalled after maintenance, the pivot bolt could unscrew by itself within a limited period of time, resulting in a hanger bar detachment.
Root causeEmployee error
ActionArjo, Inc. sent an Urgent Device Recall Customer Notification letter dated 10/29/08 to all affected end users to the attention of the Administrator/Risk Manager via UPS 2nd day on the 10/30/08. The letters informed the accounts of the potential for the hanger bar to detach from the scale if the spring pin is not replaced into the pivot bolt after performing maintenance on the lift. The accounts were advised to review the enclosed Urgent Field Safety Notice, note the affected model and serial numbers, and inspect their lifts for the presence of the spring pin in the pivot bolt. If the user fin
Quantity44 lifts, 2 scale kits
DistributionNationwide Distribution.