Recall Z-0817-2009 ยท DSI Detector And Alarm, Arrhythmia

Philips Medical Systems

Philips M8000-64100 Table Mount Shipped standard with all IntelliVue Patient Monitors MP40-70, Gas Modules Gl and G5 and certain options of the IntelliVue Patient Monitors MP5 MPST and MP20/30. Product shipped between 6/03/08 and 9/12/08 as follows: M1013A IntelliVue G1 Gas Module M1019A IntelliVue GS Gas Module M8001A option E22 Quick release mount for IntelliVue MP20 M8002A option E22 Quick r

DateJan 13, 2009 (initiated Oct 22, 2008)
StatusTerminated
ReasonTable Mounts assembled with incorrect length hardware may fall
Root causeProcess control
ActionPhilips issued an "Urgent Medical Device Correction" notification to all potentially affected customers on 10/22/08 via UPS for the US customers. The letter mailed to customers outside the US by Philips' regulatory contacts in each country via a tracked method. The letter informs customers of the problem and gives instructions on how to identify affected table mounts. Field Service Engineers will be dispatched to each customer site to inspect each Table Mount. If customers detect any sign of a loose connection between the monitor and the mounting plate while they are awaiting the Field Ser
Quantity14,292 units
DistributionNationwide Canada Foreign: Argentina, Aruba, Australia, Austria, Bahain, Bangladesh, Belgium, Bosnia, Brazil, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Egypt, Finland, France, Gabon, Germany, Great Britain, Greece, Hong Kong, Iceland, India, Iraq, Ireland , Italy, Japan, Korea, Liby
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082633INTELLIVUE PATIENT MONITOR, INTELLIVUE XDS, MODELS MP2, MP5, MP20, MP30, MP40, MP50, MP60, MP70, MP80, MP90