ClearanceBrief
Recall Z-1029-2009 ยท KIJ Instrument, Surgical, Orthopedic, Dc-Powered Motor And Accessory/Attachment

Stryker Instruments Div. of Stryker Corporation

Stryker Sagittal Saw, For use with Stryker System 6. Stryker Instruments, Kalamazoo, MI; REF 6208-000-000S4. The product is an oscillating cutting device used for cutting bone and bone related tissue.

DateMar 10, 2009 (initiated Oct 10, 2008)
StatusTerminated
ReasonThe handpieces may continue to run even when the trigger is not depressed, which could result in the unintended cutting of bone or soft tissue damage.
Root causeComponent design/selection
ActionConsignees were notified of the problem via letter dated October 10, 2008, provided interim instructions to reduce the risk, and informed that Stryker instruments will replace the trigger on all handpieces. Consignees were also instructed on the following: 1) Review the letter and distribute to all affected departments, 2) Complete and sign the enclosed Business Reply Form and fax to 336-499-8603 and 3) Call Stryker Instruments Service Department at 888-308-1983 to schedule service for the handpieces.
Quantity25,312 of all units
DistributionWorldwide Distribution including USA and countries of Australia, Brazil, Canada, Chile, Colombia, Costa Rica, England, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Pakistan, Panama, Peru, Philippines, Poland, Romania, Saudi Arabia, Sin
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K943323STRYKER SURGICAL INSTRUMENTS AND ACCESSORIES
Stryker Corp., Nov 28, 1994
K972367STRYKER SYSTEM 4000 HEAVY DUTY BATTERY POWERED EQUIPMENT
Stryker Instruments, Sep 4, 1997