ClearanceBrief
Recall Z-0941-2009 ยท FNL Bed, Ac-Powered Adjustable Hospital

Stryker Medical Div. of Stryker Corporation

Stryker InTouch 1.0 Zoom Critical Care Bed; Model 2140. The device is intended to be used as an Intensive Care Unit bed.

DateFeb 25, 2009 (initiated Jan 5, 2009)
StatusTerminated
ReasonThe touch screen on the footboard may display a "tilt over range" error, resulting in a precautionary stop of the bed motions which result in the bed appearing to have locked out some of its bed positioning features.
Root causeDevice Design
ActionConsignees were notified of the problems via letter dated January 5, 2009, advised to check their beds and to take them out of service if the foot board touchscreen displayed a "tilt over range" error, and that a Stryker Medical field service representative would contact them to perform software upgrades on the affected beds. Please direct questions to Stryker Medical at 1-269-324-6644.
Quantity1,288 of all models
DistributionWorldwide Distribution including USA and countries of Canada, Japan and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.