Recall Z-0444-2009 ยท DYB Introducer, Catheter

Medtronic, Inc.

Medtronic Input TS 7F 11 cm Ref: 071102A Percutaneous catheter introducer sets, including sheath, dilator, guide wire and needle, are used to facilitate placing a catheter through the skin into a vein or artery. Percutaneous introducers are recommended for initial percutaneous introduction or the exchange of intravascular devices. Guide wire introducer needles are used for percutaneous punctur

DateDec 2, 2008 (initiated Oct 7, 2008)
StatusTerminated
ReasonFailure to insert the guidewire through the 18 gauge introducer needle
Root causeNonconforming Material/Component
ActionMedtronic notified Medtronic Sales representatives via UPS for hand delivery of the Urgent Notification of Product Recall letter to the affected customers on 10/7/08 advising users to segregate the product so that it would not be used. The letter stated that a Medtronic Sales Rep would visit the site to account for all units that were received and would initiate a credit, as appropriate. Customers were requested to cooperate and assist with completing the necessary documentation.
Quantity335 units
DistributionWorldwide Distribution --- USA including states of IL and FL, and countries of Belgium, Honduras, and Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K955785TS PERCUTANEOUS ARTERIAL/VENUS CATHETER INTRO SET
C.R. Bard, Inc., Mar 12, 1996