Recall Z-0320-2009 ยท JAK System, X-Ray, Tomography, Computed

Siemens Medical Solutions USA, Inc

Somatom Sensation 40 X-Ray System, Model number 8874427. This is a computed tomography x-ray system which is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles.

DateJan 5, 2009 (initiated Sep 3, 2008)
StatusTerminated
Reasonsoftware error - does not allow protocol to properly synchronize with ECG triggers
Root causeSoftware design
ActionThe recalling firm issued a Customer Safety Advisory Notice dated 9/3/08 to affected customers via Update Instructions CT045/08/S. This letter informs customers of the potential issued and provides instruction to avoid its occurrence. For additional information, contact Siemens Medical Solutions USA, Inc. at 1-610-448-4500.
Quantity168 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040665SOMATOM SENSATION 64 AND SENSATION CARDIAC
Siemens Medical Solutions USA, Inc., Apr 2, 2004