ClearanceBrief
Recall Z-0265-2009 ยท EBF Material, Tooth Shade, Resin

Kerr Corp

PREMISE, PART NO. 32823, 10 PACK UNIDOSE PREMISA C4 (INTERNATIONAL ONLY), dental composite

DateNov 25, 2008 (initiated Sep 5, 2008)
StatusTerminated
ReasonMaterial stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
Root causeComponent design/selection
ActionKerr sent an Urgent Medical Device Recall letter to consignees via USPS first class mail the week of September 8, 2008, requesting they return the product for replacement or credit. The recall notification requested the return of the affected products because the material appears to stiffen and become difficult to extrude over time. Customers were instructed if they have any of the affected lots of product in their inventory to immediately return it to Kerr. Any product returned will be replaced at no charge or credited to your account. In addition, customers were instructed to fax the attac
Quantity22,478 packs for all affected lots
DistributionWorldwide: USA, Australia, Canada, Hong Kong, Malaysia, Mexico, and Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060472PREMISE FLOWABLE
Kerr Corporation, Apr 28, 2006