Recall Z-0484-2009 ยท MLR Sterilizer, Chemical

Advanced Sterilization Products

STERRAD NX Sterilizer, Product Code 10033 Low temperature sterilizers that use hydrogen peroxide gas plasma to inactivate microorganisms on a broad range of medical and surgical instruments.

DateDec 2, 2008 (initiated Oct 7, 2008)
StatusTerminated
ReasonASP has discovered a component defect in some of the UV lamp power supplies used in certain STERRAD NX Sterilizers. This defect can potentially cause the Hydrogen Peroxide Monitor to give inaccurate readings.
Root causeNonconforming Material/Component
ActionThe Urgent: Product Correction customer notifications, dated October 6, 2008, were hand delivered by field representatives starting October 7, 2008. The letter advises that Advanced Sterilization Products has discovered a component defect in some of the UV lamp power supplies used in certain STERRAD Sterilizers. This defect can potentially cause the Hydrogen Peroxide Monitor to give inaccurate readings. A list of the affected sterilizer serial numbers is included with the letter. The letter advises that the Field Service Engineer has come to their facility to inspect and, if necessary, repl
Quantity32 potentially affected systems in the US - 20 installed, 5 uninstalled, 7 potentially affected. There are also 15 units
DistributionWorldwide Distribution --- USA including states of WV, CA, TN, AZ, WI, NC, MO, NJ, DE, GA, SC, WA, FL, MA, NM, MI, and AL and countries of Ecuador, China, Israel, Japan, Korea, Middle East, Australia, Czech Republic, Spain, Brazil, and Russia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042116STERRAD NX STERILIZER
Advanced Sterilization Products, Apr 22, 2005