Recall Z-0449-2009 ยท KGI Densitometer, Bone

Hologic, Inc.

Hologic, Inc., 10 Year Fracture Risk Questionnaire Option for QDR X-Ray Bone Densitometers Software

DateDec 29, 2008 (initiated Sep 24, 2008)
StatusTerminated
ReasonSoftware error may lead to a high estimate of major fracture probability.
Root causeSoftware design
ActionHologic notified customers by telephone on 9/23 and 24/08 and instructed them to not use the software. A customer notification letter was sent via FedEx to customers on September 25, 2008.The letter requested de-installation of the software and return of the software to Hologic in a pre-paid envelope.
Quantity15 software discs
DistributionNationwide distribution in: CO, MA, MN, NM, NY, NC, OR, OH, PA, and TX. Also distributed in the UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K0807111.10-YEAR FRACTURE RISK QUESTIONNAIRE OPTION FOR QDR X-RAY BONE DENSITOMETER, MODEL QDR OPTION
Hologic, Inc., Sep 2, 2008