Recall Z-0581-2009 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Dade Behring, Inc.

Dimension EXL Clinical Chemistry System used with software versions 8.5.1 and 8.5.1SP3. The system is a discrete, random-access, microprocessor-controlled, integrated instrument/chemistry system that measures a variety of analytes, including enzyme activities, in body fluids for in vitro diagnostic use.

DateJan 13, 2009 (initiated Sep 30, 2008)
StatusTerminated
ReasonError messages, "Wash Aspirator Failure" and "Wash station unable to recover" are not generated when one or both HM wash probes are totally occluded.
Root causeSoftware design
ActionThe recalling firm notified consignees by telephone to inform them of the problem. A follow-up letter, "Urgent Field Safety Notice" dated 9/30/08 was issued with information describing the problem, the possible impact of an occluded wash probe on assay results, and emphasizing regular performance of routine wash probe maintenance as detailed in the operator guide to prevent wash probe occlusion. For questions, please contact the Siemens Technical Solution Center at 800-441-9250.
Quantity57 units
DistributionNationwide Distribution including states of AR, CA, CO, IA, ID, IL, IN, LA, MD, MI, MO, MS, NC, NE, OH, OK, PA, SD, TN, TX, VA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K073604DIMENSION EXL WITH LM CLINICAL CHEMISTRY SYSTEM, DIMENSION FT4L FLEX REAGENT CARTRIDGE, AND LOCI THYROID CALIBRATOR, MOD