Recall Z-0592-2009 ยท LKK Pump, Infusion, Implanted, Programmable
Medtronic Neuromodulation
Medtronic SynchroMed II Programmable Pump, Model 8637-40. 40 mL Reservoir. The contents [STERILE] of the inner package have been sterilized by ethylene oxide gas. The implantable Medtronic SynchroMed II programmable Pump is part of the SynchroMed II Infusion System designed to contain and administer prescribed drugs to a specific site. The implantable components of the SynchroMed II Infusion S
| Date | Jan 26, 2009 (initiated Aug 25, 2008) |
| Status | Terminated |
| Reason | Medtronic issued a letter to healthcare professionals providing important safety information regarding MRI (magnetic resonance imaging) effects on SynchroMed EL and SynchroMed II implantable infusion pumps. As stated in product labeling, the magnetic field of an MRI will temporarily stop the rotor of the pump motor and suspend drug infusion for the duration of MRI exposure for all SynchroMed pump |
| Root cause | Labeling Change Control |
| Action | An Urgent Medical Device Correction, Important Information on Potential MRI Effects letter, dated August 2008, was sent to health care professionals beginning August 25, 2008. The letter described the issue, provided patient management recommendations and included enclosures: Roller Study Procedure to confirm pump function, Post MRI Pump Interrogation Guidelines and New Information Regarding Potential MRI Effects. |
| Quantity | Approx 76,253 of all devices. |
| Distribution | Worldwide Distribution --- USA including Washington D.C., and Puerto Rico, and countries of Algeria, Aruba, Austria, Belarus, Belgium, Croatia, Cyprus Czech Republic, Denmark, Egypt, Finland, France, French Polynesia, Germany, Greece, Hungary, Iceland, India, Iran, Iraq, Ireland, Israel, Italy , Jor |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.