Recall Z-0339-2009 ยท IYE Accelerator, Linear, Medical

TomoTherapy Incorporated

TomoTherapy HI-ART Systems with 3.1.2, 3.1.3 or 3.2.1 software. The affected applications include Planning Station, Planned Adaptive, Data Management System, and TomoPortal. The product is intended for use in radiation therapy.

DateJan 5, 2009 (initiated Sep 18, 2008)
StatusTerminated
ReasonTomoTherapy Inc. discovered an issue on the Hi-Art system during the course of ongoing testing. It has been identified that treatment plans involving very small structures (volume <2cc), reading in a high dose gradient region, may exhibit inaccuracies with the DVH curve and the Dose Statistics for the small structures. This error can only happen if planning was performed using very small voxels
Root causeSoftware design
ActionConsignees were sent on 9/18/08 a "TomoTherapy Incorporated Urgent Medical Device Correction Field Safety Notice" letter dated September 17, 2008. The letter described the issue, product affected, recommended actions and resolution. Questions may be directed to TomoTherapy Inc. Customer Service Center at 1-868-368-4807.
Quantity165
DistributionAK AL AR AZ CA CO CT DC FL HI IA IL IN KY LA MA MD MI MN MO NE NJ NM NV NY OH OK OR PA PR SC SD TN TX VA WA WI and WY Belgium Canada Switzerland Germany Spain France Great Britain India Italy Japan Korea Netherlands Turkey and Taiwan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060912MODIFICATION TO: HI-ART SYSTEM
Tomotherapy Incorporated, Apr 19, 2006