ClearanceBrief
Recall Z-0434-2009 ยท KNT Tubes, Gastrointestinal (And Accessories)

C R Bard Inc

Dual Port Wizard Low Profile Replacement Gastrostomy Device, Code 00221W, 18F, 1.2cm, 10cc, Sterile EO, Bard Access Systems, Inc., Salt Lake City, UT 84116.

DateDec 12, 2008 (initiated Sep 23, 2008)
StatusTerminated
ReasonGastrostomy device anti-reflux valve may allow leakage from the stomach.
Root causeComponent design/selection
ActionCustomers notified by letter (Product Field Action letter) on September 23, 2008 and instructed to remove and return any affected product to Bard Access Systems, Inc (BAS). Customers were instructed in the recall letter to respond by completing and faxing the enclosed Reply Form and Inventory Reconciliation Form. For assistance, contact BAS customer service at 800-290-1689.
Quantity309 units
DistributionWorldwide Distribution including US, Latin America and countries of Canada, Japan, Finland, Norway, Sweden, Austria, Belgium, Holland, Great Britain, Italy, France, Switzerland, Spain, Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013144DUAL PORT WIZARD LOW PROFILE REPLACEMENT GASTROSTOMY DEVICE; DUAL PORT WIZARD FEEDING/DECOMPRESSION TUBE, MODEL 00220W
C.R. Bard, Inc., Oct 19, 2001