ClearanceBrief
Recall Z-0120-2009 ยท IZI System, X-Ray, Angiographic

Siemens Medical Solutions USA, Inc

Axiom Artis dFC, Axiom Artis Modular Angiography X-Ray System, Model number 7412807.

DateNov 25, 2008 (initiated Sep 2, 2008)
StatusTerminated
ReasonIncorrect contour finding: Image may be calibrated to the wrong size catheter.
Root causeSoftware design
ActionSiemens issued a Customer Safety Advisory Notice dated 8/29/08 to inform customers of the potential issue. The recalling firm issued Update Instructions AX040/08/S to the service engineers which provides information to fix the problem.
Quantity145 units
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010721AXIOM ARTIS MODULAR ANGIOGRAPHY SYSTEM
Siemens Medical Corp., Mar 30, 2001
K021021MODULAR ANGIOGRAPHY SYSTEM AXIOM ARTIS
Siemens Medical Solutions USA, Inc., Jun 6, 2002