Recall Z-0200-2009 ยท IZO Generator, High-Voltage, X-Ray, Diagnostic

GE Healthcare

Monitor suspensions used with Advantx Legacy, RFX Classical, SFX-90, Model # 46-240838G2, manufactured after March 8, 1985 for Advantx Legacy, RFX and SFX systems. General-purpose radiographic system component used to suspend the system's image monitor.

DateOct 27, 2008 (initiated Mar 9, 2007)
StatusTerminated
ReasonGE Healthcare has recently become aware of a potential risk of monitor dropping associated with the monitor suspension of the Advantx system that may impact patient safety. It has been reported that suspensions manufactured after March 8, 1995 did not have the required thread locking agent applied to the setscrew, allowing the setscrew to back out over time, which can lead to the monitor dropping
Root causeComponent change control
ActionConsignees were sent a GE Healthcare "Product Safety Notification" letter dated September 13, 2007. The letter was addressed to Hospital Administrators/Managers of Radiology/Radiologists. The letter described the Safety Issue, Affected Product, Safety Instructions, Corrective Action and Contact Information. The letter recommends that customers move the monitor suspension to an open area in the room where viewing is still possible and make sure to minimize the movement of the monitor until a Qualified GE Healthcare Field Engineer has inspected the suspension. The letter also states that a Q
Quantity21
DistributionWorldwide Distribution --- USA including states of AZ, CA, FL, GA, IA, MD, MO, NC, NY, OH, OK, TN, TX,WI, and LA, and country of Slovak Republic.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K760466TABLE, X-RAY, RFX CLASSICAL
General Electric Co., Aug 26, 1976
K801320SFX-90 TABLE/SPOTFILMER W/FLUORICON L200
General Electric Co., Jul 21, 1980
K931049ADVANTX SYSTEM
GE Medical Systems, May 19, 1993