Recall Z-0202-2009 ยท CBK Ventilator, Continuous, Facility Use

Draeger Medical, Inc.

Oxylog 3000 Emergency and Transport Ventilator Continuous Respirator with software version 1.10 pre-installed; Catalog number 2M86300.

DateOct 27, 2008 (initiated Oct 3, 2008)
StatusTerminated
ReasonMay experience an interruption of ventilation for approximately 5 seconds
Root causeSoftware design
ActionThe firm issued an Urgent - Medical Device Recall letter dated September 2008. Software version 1.10 may under specific conditions experience an interruption of ventilation for approximately 5 seconds. During that time period, the patient airway pressure may drop to zero. This phenomenon was identified by Draeger engineers and to date has not been reported by any user. The firm's investigations revealed that when ventilating in BIPAP mode with the pressure rise time (Ramp) set to the highest setting and the expiration time set to exactly 0.5 seconds, the inspiration phase can coincide with th
Quantity50 units
DistributionNationwide Distribution --- including states of AZ, CA, FL, IL, IN, KY, LA, ME, MI, MO, NC, NE, NJ, NY, OH, PA, TX, UT, VT, WA, WV, and WY.