ClearanceBrief
Recall Z-0317-2009 ยท IZI System, X-Ray, Angiographic

Siemens Medical Solutions USA, Inc

AXIOM Artis System dMP using software version VB31D with 30x40 cm2 flat detectors. AXIOM Artis is a angiography system developed for single and biplane diagnostic imaging and interventional procedures.

DateDec 9, 2008 (initiated Aug 25, 2008)
StatusTerminated
ReasonSystem may switch to emergency fluoroscopy mode.
Root causeSoftware design
ActionA Customer Safety Advisory Letter dated 8/20/08 was issued with Update Instructions AX045/08/S. The letter informed the customer of the potential issue and provided instruction to avoid its occurrence. A software update (VB31E) was released to correct the issue.
Quantity3 units
DistributionNationwide Distribution including states of AR, CA, CO, FL, IA, IL, IN, KS, KY, MI, NY, TX, VA, and WY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052202AXIOM ARTIS - MODULAR ANGIOGRAPHIC SYSTEM
Siemens Medical Solutions USA, Inc., Mar 7, 2006