Recall Z-0163-2009 ยท KPR System, X-Ray, Stationary

Siemens Medical Solutions USA, Inc

Axiom Aristos FX Multipurpose Radiography System, stationary X-Ray system, Model number: 7414803.

DateOct 20, 2008 (initiated Aug 25, 2008)
StatusTerminated
ReasonTube support arm or the detecting mounting hardware may become loose at the joint with the telescope.
Root causeEquipment maintenance
ActionUpdated Instructions were issued in August 2008 (AX049/08/S) to Siemens Service Engineers to have them inspect and tighten any loose hardware on the affected systems.
Quantity7 units
DistributionNationwide distribution including states of CA, IN, KY, UT, VA and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013826AXIOM ARISTOS FX (MULTI PURPOSE RADIOGRAPHY SYSTEM)
Siemens Medical Systems, Inc., Feb 1, 2002