ClearanceBrief
Recall Z-0307-2009 ยท HRS Plate, Fixation, Bone

Integra LifeSciences Corp

Uni-CP Compression Forceps (spreader); Non-sterile, R only, Catalog number: 339001ND Newdeal, Lyon, France; Integra, 411 Enterprise Drive, Plainsboro, NJ 08536 Intended to open the UNI-Compression Plate olive, which creates a compression.

DateDec 1, 2008 (initiated Sep 30, 2008)
StatusTerminated
ReasonReports of the Uni-CP Compression Forceps breaking during the Compression of the Uni-CP plate
Root causeNonconforming Material/Component
ActionRecall Notice letters were sent to all Integra Sales Specialists on September 30, 2008 by first class mail. The letter informed them of the product and the problem concerning the product. The Sales Specialists were asked to return all inventories under their control with part number 339001ND from lots E6WS, E7UY, & E86M. They are required to fill out the attached form identifying the lot number for all Newdeal Compression Forceps (Spreader) in their inventory along with the serial number(s) of their Uni-CP set(s). The form should be filled out even if the sales specialist does not have a U
Quantity90
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070447NEWDEAL COMPRESSION PLATE
Newdeal Sas, Mar 29, 2007