Recall Z-0130-2009 ยท DYB Introducer, Catheter

Oscor, Inc.

Oscor Adelante Luer-Lock Peel Away Introducer Set, for introduction of diagnostic or therapeutic devices into the body.

DateOct 29, 2008 (initiated Jul 3, 2008)
StatusTerminated
ReasonDifficulty breaking the sheath hub and subsequently to peel the sheath off for Introducer Set, Adelante size 7F.
Root causeProcess design
ActionOscor notified the Direct Account by letter on 07/03/2008. Direct account was asked to return all unused Introducers back to Oscor. All returned products should be labeled with the Return Goods Product number.
Quantity396
DistributionProduct was distributed to one direct account in MN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K073100INTRODUCER SETS, MODELS ADELANTE AND ADELANTE-S
Oscor, Inc., Nov 30, 2007