Recall Z-0476-2009 ยท GCJ Laparoscope, General & Plastic Surgery

CTS, Inc. dba Guidant Cardiac Surgery

VASOVIEW 4 Endoscopic Vessel Harvesting System, with VASOVIEW Flexible Endoscopic Tools in the following configurations: VH-10242: Uniport plus dissection cannula, short port blunt tip trocar, flexible endoscopic scissors; and VH-10243: Uniport plus dissection cannula, short port blunt tip trocar, flexible endoscopic bisector.

DateDec 23, 2008 (initiated Sep 9, 2008)
StatusTerminated
ReasonPackaging unsealed: Sterility of product may be compromised due to packaging issue. For further information, please contact the firm at 408-635-6874.
Root causePackaging process control
ActionThe firm issued a letter entitled "URGENT FIELD ACTION - IMMEDIATE ACTION REQUIRED *** VASOVIEW(TM) 4 Endoscopic Vessel Harvesting System (Kit Only)" via federal express to its consignees on September 9, 2008. The firm has asked its consignees to return the product. Follow-up calls were made to ensure recall effectiveness.
Quantity7,935 units
DistributionWorldwide Distribution --- including USA and country of Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.