Recall Z-1707-2008 ยท JAK System, X-Ray, Tomography, Computed

Toshiba American Med Sys Inc

Toshiba America Medical Systems (TAMS) Aquilion 32 CT system, computed tomography x-ray system.

DateSep 29, 2008 (initiated Oct 1, 2007)
StatusTerminated
ReasonExcessive radiation: Intermittent shifting of CT numbers causes the actual dose to the patient to be greater than expected, when Sure Exposure option is used.
Root causeDevice Design
ActionToshiba sent a letter to customers in July 2007, and sent field service representatives to update the Data Aquisition System in September/October 2007.
Quantity1203 total
DistributionWorldwide distribution, including USA and Japan.
FDA recordOpen on fda.gov