Recall Z-1522-2008 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Datex - Ohmeda, Inc

GE Healthcare Innova 4100 Digital Fluoroscopic Imaging (Cardiovascular Imaging System

DateSep 30, 2008 (initiated May 19, 2008)
StatusTerminated
ReasonGE Healthcare has become aware of a potential condition that may impact operator or patient safety. One customer has reported that inability to terminate X-ray exposure after releasing the hand switch control (located in the control room).
Root causeSoftware design
ActionAn Urgent Medical Device Correction letter, dated June 2008 was sent to customers The letter described the issue, affected products and provided short term user recommendations. GE developed a new software release that corrects this condition and will launch the corrective action to upgrade all affected systems.
Quantity87
DistributionWorldwide Distribution - AL, AZ, CA, FL,, GA, HI, IA,IL, IN, KY, LA,MA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WI, and WV, Puerto Rico. OUS to include: Australia, Byelarus, China, Denmark, France, Germany, Greece, Honduras, Iran, Iraq, Italy, Japan, Latvia, Lebano