Recall Z-1521-2008 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Datex - Ohmeda, Inc

GE Healthcare Innova 3100 Digital Fluoroscopic Imaging (Cardiovascular Imaging System)

DateSep 30, 2008 (initiated May 19, 2008)
StatusTerminated
ReasonGE Healthcare has become aware of a potential condition that may impact operator or patient safety. One customer has reported that inability to terminate X-ray exposure after releasing the hand switch control (located in the control room).
Root causeSoftware design
ActionAn Urgent Medical Device Correction letter, dated June 2008 was sent to customers The letter described the issue, affected products and provided short term user recommendations. GE developed a new software release that corrects this condition and will launch the corrective action to upgrade all affected systems.
Quantity107
DistributionWorldwide Distribution - AL, AZ, CA, FL,, GA, HI, IA,IL, IN, KY, LA,MA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WI, and WV, Puerto Rico. OUS to include: Australia, Byelarus, China, Denmark, France, Germany, Greece, Honduras, Iran, Iraq, Italy, Japan, Latvia, Lebano
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031637INNOVA 3100
GE Medical Systems, Aug 22, 2003
K042053INNOVA 4100, INNOVA 3100
Ge Medical Systems, Inc., Aug 26, 2004