ClearanceBrief
Recall Z-0368-2008 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

OEC Medical Systems, Inc

GE Healthcare OEC 9900 mobile fluoroscopic x-ray system

DateSep 29, 2008 (initiated Dec 14, 2007)
StatusTerminated
ReasonPossible overexposure: Failure to apply the entrance exposure rate (EER) tube current limit calibration to the automatic exposure rate (AER) control system when the anatomical profile mode is changed from the default selection to another selection. Manual technique modes are not affected.
Root causeSoftware design
ActionGE OEC sent a letter to customers December 15, 2007, and scheduled field representative visits for corrective software solution installation.
Quantity710
DistributionWorldwide distribution: USA, Finland, Netherlands, Germany, Switzerland, France, United Kingdom, Australia, Italy, Ireland, Spain, Sweden, Belgium, New Zealand, China, Singapore, Poland, Norway, and India.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.