Recall Z-1713-2008 ยท KPR System, X-Ray, Stationary

GE Healthcare

GE Healthcare Proteus XR/a Eclipse Collimator model # 2379827.

DateSep 30, 2008 (initiated Mar 29, 2007)
StatusTerminated
ReasonThe actual average illuminance for the collimators are approximately 140-lux. This does not meet the 160-lux requirement. GE will be correcting the units by replacing all collimators.
Root causeRadiation Control for Health and Safety Act
ActionGE will be correcting the units by replacing all collimators that don't meet the 160-lux requirement. The field correction will be initiated and is scheduled to be completed by October 2008.
Quantity132
DistributionWorldwide Distribution --- USA including states of CA, DE, FL, GA, ID, IL, KS, KY, MD, MI, MO, MT, NC, NJ, NM, NY, OH, PA, SC, TX, UT, VA, and WV, and countries of Australia, Canada, China, France, Germany, Greece, Italy, Korea, Poland, Saudi Arabia, United Kingdom, and Venezuela.