Recall Z-0194-2009 ยท IZI System, X-Ray, Angiographic

Siemens Medical Solutions USA, Inc

Axiom Artis Zee. Angiographic X-ray System.

DateNov 4, 2008 (initiated Aug 22, 2008)
StatusTerminated
ReasonPatients could possibly fall off the table if not properly secured when: 1) moving them onto the patient table, 2) moving them on the patient table or 3) removing them from the patient table.
Root causeLabeling design
ActionThe recalling firm issued a Customer Safety Advisory Notice dated 7/04/08 (including an Addendum to the Operators Manual) which alerts the operator of possible injury to patients if they are not properly secured to the table during examinations and describes the recommended fastening procedures required to properly secure the patient to the table. For questions or additional information, contact Siemens Medical Solutions, Inc. at 610-448-4634.
Quantity108 units
DistributionNationwide Distribution.