ClearanceBrief
Recall Z-0407-2009 ยท GAZ Tubing, Noninvasive

Mentor Texas, Inc

Mentor Aseptic Transfer Set, Cat. #350-8400 (contains BD 60mL Luer-Lok Syringe). Aseptic transfer set manufactured by Mentor Texas, LP, Irving, TX 75038 (BD 60 mL Luer-Lok syringe manufactured by Becton Dickinson, Franklin Lakes, NJ 07417).

DateDec 12, 2008 (initiated Aug 22, 2008)
StatusTerminated
ReasonThe Mentor Aseptic Transfer Set contains a component, the BD 60mL Luer-Lok Syringe, which is under recall by Becton Dickinson due to a package integrity issue.
Root causePackage design/selection
ActionFirm sent notification, along with a copy of the Becton Dickinson recall letter, to consignees on 08/18/08. Consignees asked to inspect inventory of Mentor Aseptic Transfer Sets for affected BD 60mL syringes and to complete and return the included Device Accountability Worksheet. For additional information, please contact Mentor Customer Service at 800-235-5731.
Quantity32,480 Aseptic Transfer Sets (4 syringes/set)
DistributionWorldwide Distribution; product sent to all states (except Alaska), Washington DC and Puerto Rico. Product also sent to the following countries: Australia, Canada, Costa Rica, Korea, New Zealand and South Africa.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.