Recall Z-0182-2009 ยท FQO Table, Operating-Room, Ac-Powered

Stryker Communications Corp

Circlip component of the Stryker Flat Panel and Navigation Arm System. Stryker Flat Panel and Navigation Arm System are intended for use as a ceiling-mounted device to support a medical grade flat panel monitor or camera in the patient area. The Circlip within the flat panel and navigation arms is part of the mechanical stop designed to prevent the arm from rotating infinitely. It also prevents t

DateOct 22, 2008 (initiated Sep 5, 2008)
StatusTerminated
ReasonCirclip component used to suspend flat panel and navigation arm system may become dislodged.
Root causeDevice Design
ActionNotification letter was sent to consignees September 5, 2008. Field service will be performed at all affected sites. Old Circlips will be replaced with new clamps that do not allow the arm to fall. Contact Stryker Communications Corp. at 1-972-410-7310 for assistance.
Quantity705 units
DistributionNationwide and International Distribution