Recall Z-0569-2009 ยท GFA Blade, Saw, General & Plastic Surgery, Surgical

Sterilmed Inc

Saw Blade Model #'s: BIO506112 BIO506113 DEP2555-91-000 DEP5627-11-500 HAL5023-118 HAL5059-32 HAL5071-133 HAL5071-145 HAL5071-163 HAL5071-181 HAL5071-572 HOWHB-NST-6828-7-791 HOWHB-NST-6828-7-792 KOMK-2000-65 KOMKM-004 KOMKM-005 KOMKM-101 KOMKM-101C KOMKM-102 KOMKM-234 KOMKM-275R KOMKM-3101 KOMKM-3107 KOMKM3-225 KOMKM-325R KOMKM-619 KOMKM-660 KOMKM-693 LIN5023-170 MICSO-601 MICSO-610 MICSP-414A M

DateJan 13, 2009 (initiated Aug 8, 2008)
StatusTerminated
ReasonSteriMed, Inc is conducting a product removal of a small group of devices individually packaged and sealed within a pouch. The removal is being conducted because the possibility exists that some of the pouches were not properly sealed on one end. A breach in packaging seal or a failure in packaging integrity has the possibility of risk to the patient in terms of transmitting organisms capable
Root causePackaging process control
ActionConsignees were notified of the recall via verbal communication in person by on-site technician and or their local sales representative on 8/8/08. Others were contacted via fax or e-mail followed by on-site visits 8/11/08. SteriMed also provided consignees (Materials Manager/Risk Manager) with a "Urgent Product Removal" recall letter (August 14, 2008) in person informing them of the potential of receiving affected product. Included in the Product Removal Notification was a Customer Reply Form to inventory quantity of product received and quantity returned. Contact SterilMed, Inc. at 1-888-
Quantity370
DistributionAL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WI, and WV OUS: CANADA