Recall Z-0550-2009 ยท GFF Bur, Surgical, General & Plastic Surgery

Sterilmed Inc

Burrs Model #'s: ANSL-3SB ANSL-5D ANSQD11-4OVDC HAL5056-118 HAL5056-166 HAL5091-106 HAL5091-112 HAL5091-122 HAL5091-125 HAL5091-136 HAL5091-144 HAL5091-164 HAL5092-120 HAL5092-168 HAL5092-170 KOM04-D0139 LIN5091-101 LIN5091-102 LIN5091-142 LIN5092-132 LIN5092-160 MED874-447 MICZB-228 MICZB-232 MID14CY50 MID14MH30D MIDG6-450 MIDG8-130 MIDMC254 STR1607-2-35 STR1608-6-137 STR1608-6-141 STR1608-6-14

DateJan 13, 2009 (initiated Aug 8, 2008)
StatusTerminated
ReasonSteriMed, Inc is conducting a product removal of a small group of devices individually packaged and sealed within a pouch. The removal is being conducted because the possibility exists that some of the pouches were not properly sealed on one end. A breach in packaging seal or a failure in packaging integrity has the possibility of risk to the patient in terms of transmitting organisms capable
Root causePackaging process control
ActionConsignees were notified of the recall via verbal communication in person by on-site technician and or their local sales representative on 8/8/08. Others were contacted via fax or e-mail followed by on-site visits 8/11/08. SteriMed also provided consignees (Materials Manager/Risk Manager) with a "Urgent Product Removal" recall letter (August 14, 2008) in person informing them of the potential of receiving affected product. Included in the Product Removal Notification was a Customer Reply Form to inventory quantity of product received and quantity returned. Contact SterilMed, Inc. at 1-888-
Quantity125
DistributionAL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WI, and WV OUS: CANADA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.