Recall Z-1882-2008 ยท KPR System, X-Ray, Stationary

GE Healthcare

GE Healthcare Definium 8000 Digital Radiographic System.

DateSep 25, 2008 (initiated Mar 7, 2008)
StatusTerminated
ReasonRadiation dose reading greater than actual dose: Error in the generator software on the GE Healthcare Revolution XR/d configured with HP 8200 PC consoles and all Definium 8000 systems may impact patient safety. Occasional generator software errors may cause light X-ray Images with reported mAs readings higher than was actually exposed to the patient using the overhead X-ray tube. The mAs rea
Root causeSoftware design
ActionGE sent a Product Safety Notification dated March 7, 2008 to all customers. The letter described the safety issue, affected product, and stated that a GE Healthcare Field Engineer will schedule a field visit to install software updates.
Quantity454 (277 US / 177 OUS)
DistributionWorldwide Distribution: USA, Australia, Austria, Belgium, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Guyana, Hong Kong, Ireland, Israel, Italy, Japan, Kuwait, Martinique (French), Netherlands, Norway, Palestine, Portugal, Saudi Arabia, South Korea, Spain, Sweden
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051967REVOLUTION XR/D DIGITAL RADIOGRAPHIC IMAGING SYSTEM WITH TOMOSYNTHESIS
GE Healthcare, Aug 9, 2005