Recall Z-0102-2009 ยท LLZ System, Image Processing, Radiological

Siemens Medical Solutions USA, Inc

Syngo Imaging with versions VB20B, VB20D, VB20F, VB20G, VB25B, VB30A, VB30A_SP1, and VB30A_SP2.

DateOct 15, 2008 (initiated Aug 22, 2008)
StatusTerminated
ReasonImage may not be visible after merge.
Root causeSoftware design
ActionThe recalling firm issued Customer Safety Advisory Notice dated 8/22/08 to inform the consignees of the potential issue and provide instruction to avoid its occurrence. To avoid potential risk of this issue, users are asked to check whether the extended HL7 messages as stated in the letter are mandatory for their RIS-PACS workflow, and disable these messages if they are not. Siemens is preparing a modification under the UI IM086/08/S for the affected versions of syngo Imaging that will resolve the potential malfunction. The field modification will be immediately distributed and installed by
Quantity59 units
DistributionNationwide Distribution --- including states of AL, AZ, CA, CO, CT, FL, IA, KS, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, TN, TX, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052461SYNGO IMAGING VERSION V20A
Siemens AG Medical Solutions, Sep 22, 2005
K071114SYNGO IMAGING (VERSION V30)
Siemens AG Medical Solutions, May 22, 2007