Recall Z-0034-2009 ยท GAR Suture, Nonabsorbable, Synthetic, Polyamide

Surgical Specialties Corp

3-0 Black Mono Nylon Suture, 12 units per box. product number 925B. The product intended for use in general soft tissue approximation and/or ligation including use in ophthalmic procedures.

DateOct 2, 2008 (initiated Sep 9, 2008)
StatusTerminated
Reasonmislabeled product - labeled as 3-0 black mono nylon contains 4-0 black braided silk suture
Root causeEmployee error
ActionThe recalling firm issued a Recall letter dated 9/9/08 to its customers. The letter informed the customers of the problem, the need to conduct a sub-recall, and return the product. Contact Surgical Specialties Corp. at 1-610-404-1000 for assistance.
Quantity956 boxes
DistributionUSA--AZ, CA, CT, FL, GA, IA, IL, KS, KY, LA, MD, MI, MS, NC, NJ, NY, OH, TN, TX, VA, and WA. OUS: Australia, Canada, and Fiji Islands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K930825SUTURE NONABSORBABLE-NYLON,SILK,POLYESTER,STAINLES
Surgical Specialties, Inc., Jan 7, 1994