ClearanceBrief
Recall Z-2434-2010 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips Medical Systems North America Co. Phillips

Panorama 0.6T MRI Nuclear Magnetic Resonance Imaging System, Philips Panorama 0.6T Open MRI system, Picker International, Inc. World Headquarters 595 Miner Road Cleveland, Ohio 44143

DateSep 17, 2010 (initiated Sep 19, 2008)
StatusTerminated
ReasonWater may collect and freeze in the vent pipe. This can rupture the magnet vessel during a cryogen transfer, and release helium into the MRI suite.
Root causeOther
ActionPhilips issued an Urgent Correction letter dated 09/19/2008 to their consignees, identifying the reason for the recall and the affected products. The firm issued the Field Change Order (FCO) 78100232. Field service engineers will visit each consignee and conduct an inspection of the magnets venting system of Edge, Vista, Eclipse, Polaris, Infinion, and Panorama 0.6T MRI systems. The consignees should contact the Philips Call Center at 1-800-772-9377, #5 #2 and reference "FCO 78100232".
Quantity38 units
DistributionNationwide Distribution: USA., including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY. There is no distribution to foreign consignees.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013858INFINION 0.6T MR IMAGING SYSTEM
Philips Medical Systems (Cleveland), Inc., Jan 22, 2002