Recall Z-1883-2008 ยท KPR System, X-Ray, Stationary

GE Healthcare

GE Healthcare Definium 8000 Digital Radiographic System, model number 5135678. GE Healthcare, 3000 North Grandview, Waukesha, WI 53188 U.S.A. The intended use of this product is to generate tomographic images of human anatomy.

DateSep 25, 2008 (initiated Jan 2, 2008)
StatusTerminated
ReasonNot properly labeling the manual switch for disabling positive beam limitation (PBL) located on the overhead tube support user interface as required
Root causeRadiation Control for Health and Safety Act
ActionGE Field Engineer will visit each site to implement the correct label. Contact GE Healthcare at 1-262-544-3894 for assistance.
Quantity668 (419 US / 249 OUS)
DistributionAL, CA, CO, CT, DE, FL, GA, IA,, IL,, IN, KS, LA, MA, MD, MI, MN, MO, MT, NC, ND, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI,WV and Washington D.C. OUS to include: Australia, Austria, Belgium, China, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Guyana, Hong Ko
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051967REVOLUTION XR/D DIGITAL RADIOGRAPHIC IMAGING SYSTEM WITH TOMOSYNTHESIS
GE Healthcare, Aug 9, 2005