Recall Z-1884-2008 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

OEC Medical Systems, Inc

GE OEC 9800 Fluoroscopic X-Ray System, GE Healthcare, Surgery, Salt Lake City, Utah. The intended use of this device is to perform mobile fluoroscopic examination of human anatomy.

DateSep 25, 2008 (initiated Aug 12, 2008)
StatusTerminated
ReasonBeam limitation may be non-compliant on some X-ray units.
Root causeRadiation Control for Health and Safety Act
ActionConsignees were notified by letter on 08/12/2008. GE service representatives will check proper beam compliance on next scheduled maintenance visit or will schedule a visit to address the issue. Information available at 1-800-874-7378.
Quantity201 systems
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K024012OEC 9800 E/CV+ DIGITAL MOBILE IMAGING SYSTEM
Ge Oec Medical Systems, Mar 20, 2003