Recall Z-1709-2008 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

OEC Medical Systems, Inc

GE OEC 9600 C-arm Fluoroscopy System, GE Healthcare, Surgery, Salt Lake City, Utah.

DateSep 20, 2008 (initiated Jul 2, 2008)
StatusTerminated
ReasonSecondary collimator filters may be missing on certain x-ray units. It was discovered that the secondary collimator filter on some OEC 9600 C-arms was omitted when a collimator assembly was replaced. This would reduce the filtration and therefore the Half-Value-Layer (HVL) of the X-ray beam, and may result in non-compliance with respect to Maximum Entrance Exposure Rate (EER) unless the maximum EE
Root causeEquipment maintenance
ActionDomestic customer notification letters were sent via certified mail on 07/02/2008. They were advised to have the Maximum EER assessed and to discontinue use if the maximum limits were exceeded.
Quantity660 units
DistributionNationwide Distribution including states of CA, FL, MS, and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K963952SERIES 9600 MOBILE DIGITAL IMAGING SYSTEM
Ge Dec Medical Systems, Dec 23, 1996