ClearanceBrief
Recall Z-0068-2009 ยท FGE Stents, Drains And Dilators For The Biliary Ducts

Bard Peripheral Vascular Inc

Edwards Lifesciences LifeStent FlexStar Self-Expanding Biliary Stent System, Edwards Lifesciences LLC Irvine, CA 92614, Model EX090401CD. The LifeStent FlexStar self-expanding biliary stent system is intended for use in the palliation of malignant strictures (neoplasms) in the biliary tree.

DateOct 10, 2008 (initiated Aug 29, 2008)
StatusTerminated
ReasonSome LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary sheath such that the guidewire lumen is visible.
Root causeProcess control
ActionThe customer letters for the tip to sheath gap were sent on 8/29/2008 with return receipt. The letter advises that C.R. Bard has determined that some LifeStent FlexStar Systems may exhibit a gap between the tip of the delivery system and the primary sheath such that the guidewire lumen is visible. Per the letter, the firm's analysis has determined that a gap between the tip of the delivery system and the delivery system sheath may be associated with difficult advancement to the treatment site. Users are instructed to inspect all devices to verify that there is not a gap between the tip and
Quantity209 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K053404LIFESTENT FLEXSTART BILIARY STENT SYSTEM
Edwards Lifesciences, LLC, Dec 21, 2005