Recall Z-0027-2009 ยท GCJ Laparoscope, General & Plastic Surgery

Applied Medical Resources Corp

Applied Medical, Lap Banding Kit, REF: K0398 Contains: (2) C0Q04, 5x100mm Threaded Kii Access System (1) C0604, 15x100mm Non-Threaded Separator System w/Universal Seal, for laparoscopic procedures.

DateOct 7, 2008 (initiated Jun 25, 2008)
StatusTerminated
ReasonPotential inability to insufflate through the stopcock. Under certain circumstances, a flexible elastomeric component inside the 15mm trocar can stretch into a configuration that blocks the flow of insufflation gas (C02).
Root causeProcess control
ActionCustomers were notified via FedEx or UPS on June 25,2008 with an overnight letter. The letter advises that Applied Medical is conducting a voluntary recall of the 15mm Separator Access Systems, Models C0604, C0605, C0606 and C0607 due to the potential inability to insufflate through the stopcock. Per the letter, the firm is asking that all 15mm product listed in the letter be returned. A response form is provided.
Quantity14,113 total
DistributionNationwide and Canada distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012884DILATING TIP OBTURATOR
Applied Medical Resources Corp., Jan 17, 2002